The ground bond test (also called PE resistance test, ground continuity test), is the first test required by the electrical safety testing standards.
The ground bond test consists in testing whether the ground points of a device under test are well connected in between each other, and also to the mains ground.
The measure result of a ground bond test is a resistance value, which has to be lower than the indicated limit from the international standards.
FOUR WIRE MEASUREMENT (KELVIN METHOD) FOR GROUND BOND TEST
When measuring low value resistances (for the ground bond test for instance), an important error source is coming from the contact resistance. Actually, in many applications, the contact resistance value can go beyond the value which has to be measured.
The method to cancel this error source, is to supply a high stability permanent current over 2 wires and to combine a voltage measurement on 2 other wires. The voltmeter which is used to measure the voltage drop at the resistance terminals, must have a high input impedance and a good measurement accuracy.
The diagram opposite shows the 4 wire measurement principle used in our milliohmmeters (ground bond test).
Because of the high input impedance of the voltmeter, there is no error due to the leads resistance and the contacts in the voltage measurement circuit. Actually any current is flowing through these wires.
We are offering a wide range of 4 wire accessories (Kelvin type) to make easy the connection on the specimen to be measured.
接地連續(xù)性測試標(biāo)準(zhǔn)要求
測試目的:驗證所有與操作者有接觸的導(dǎo)電部件接地是否良好
測試方法:加載一個合適的電流,測試電壓值,得出阻抗值
一般選用四線測量法(開爾文測試),最大程度去除接觸阻抗影響(線損分流)
推薦型號:50-C 56-S 506-S
測試標(biāo)準(zhǔn) | EN61010-1 | EN60335-1 | EN60950 | EN60598-1 | EN60204-1 | EN61131-2 |
測試電壓 | 10V AC or DC | 12V | < 12V | < 12V | 無要求 | < 12V |
測試電流 | 25A AC or DC | 25A AC or DC | 25A AC or DC | ≤ 10 | ≥10A | 30A |
閾值 | 0.1? | 0.1? | 0.1? | 0.5? | 0.1? | 0.1? |
測試時間 | 1min | 無要求 | 無要求 | 無要求 | >10s | ≥ 2min |